At Scientifica we take your privacy seriously and will only use your personal information to provide the products and services you have requested from us. If you consent to us collecting and storing your data from this form please tick the checkbox below. You can withdraw your consent at any time. For more information please view our Privacy Policy *

npi electronic PDES - Pneumatic drug Ejection

PDES - Pneumatic Drug Ejection npi units are designed for pressure ejection of drugs in physiological and pharmacological studies.

npi electronic PDES - Pneumatic drug Ejection

PDES systems can be equipped with high pressure outputs (up to Pmax = 4 bar / 87 psi) for drug application with fine-tipped micropipettes or with low-range pressure output for controlled droplet application from large-tipped pipettes. Analog (standard) or digital timers (optional) and a remote control unit facilitate the use of these instruments.

Each system is housed in a 19" cabinet that can be mounted in a rack. It comprises of two channels. Each channel consists of a precision pressure regulator, an analog (or digital) pressure display, digital (or analog) timer, electronically controlled valves, control electronics and a regulated power supply. Ejection of drugs can be activated manually or via a TTL input. The operational status is indicated by a red / green LED. Each channel is equipped with an "EXHAUST" coupling whereby a low "retain" pressure can be applied (analogous to the "retaining" current used in iontophoretic systems).

PDES systems are available with valves located in the controller or very fast application, with a micro valve located in an injection pipette holder (microJECT).

Source gas
Source gas
Air, nitrogen, helium etc., no aggressive, combustible or explosive gases, filtered by a 5 μm filter element
Pressure range
Pressure range
Input pressure
Input pressure
Max 5 bar (72psi)
Pressure control range
Pressure control range
0-4 (0-58psi)
Residual air pressure
Residual air pressure
10 mbar (0.145 psi) to 100 mbar (1.45 psi)
Input and output couplings
Input and output couplings
Quick connect type
Tubes
Tubes
Input: 6mm OD
Output: 4mm OD
Mode selection
Mode selection
Three position toggle switch (TIMER, CONT., EXT.)
Minimum valve switch time
Minimum valve switch time
Standard valve: approx. 10ms dependent on the tubing length with microJECT: < 1ms
Digital timer
Digital timer
100 μs - 10 s set by five digit switch (0000.0 - 9999.9 ms)
Analog timer
Analog timer
0.001 - 0.1 s, 0.01 - 1 s, 0.1 - 10 s and 1 - 100 s linear control, range set by toggle switch time base selector (0.01 s, 0.1 s, 1 s, 10 s)
Pulse input (BNC connector)
Pulse input (BNC connector)
For external operation: TTL (LO= 0-2.5 V, HI= 3-5 V) input impedance: 10 k
Timing output (BNC connector)
Timing output (BNC connector)
TTL (LO= 0-2.5 V, HI= 3-5 V) output impedance: 249 Ω
Pressure output (BNC connector)
Pressure output (BNC connector)
Pressure monitor:, sensitivity: 1 V/bar or 0.1 V/psi, output impedance: 249 Ω
Remote control (option)
Remote control (option)
Toggle switch and status LED for each channel, size: 2.5 x 4 x 10cm
Dimensions
Dimensions
19" rackmount cabinet, 19" (483 mm), 10" (250 mm), 3.5" (88mm)